These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of mouse IL-17 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 95 | ||||
Range % | 89-101 | |||||
1:2 | Average % | 95 | ||||
Range % | 90-100 | |||||
1:4 | Average % | 88 | ||||
Range % | 81-99 | |||||
1:8 | Average % | 90 | ||||
Range % | 81-98 |
The recovery of mouse IL-17 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 93 | 88-100 | ||||
EDTA plasma (n=4) | 94 | 90-99 | ||||